7.2.2 Overwrite internal adjustment weight
(Models with settings not appropriate for verification only)
! Overwriting is restricted to specialist staff possessing well acquainted with the
workings of weighing scales.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
Press the MENU button as often as required until „CAlib“ appears
Operate the PRINT key
Press the MENU button as often as required until „TEC-CAL“ appears
Press the PRINT button and keep it pressed until the acoustic signal gets mute.
Press the MENU button and keep it pressed until the acoustic signal gets mute. The
balance changes automatically into weighing mode
Ensure that there is no load on the weighing plate
Press the CAL key
Wait until the exact value of the adjustment weight will be shown flashing
(Example)
Place the displayed adjusting weight on the weighing plate.
The flashing display extinguishes and the balance changes into weighing mode.
Remove adjustment weight from weighing plate
Press the PRINT button and keep it pressed until the acoustic signal gets mute.
The calibrating process is started
During this processtEc MEM“ is displayed
After having automatically saved the value of the internal adjustment weight, the balance
will return into weighing mode.
Carry out the adjustment process as described in chapter 7.2