24 MLS_C-BA-e-1330
Place the adjustment weight in the centre of the
sample dish, then press the PRINT key. If the
adjustment was successful, this will be followed by
automatic reset to the menu.
The ESC key may be used to cancel the adjustment.
Errors during adjustment or the use of an incorrect
adjusting weight will result in an error message.
Repeat adjustment.
Return to weighing mode:
Press the ESC-key repeatedly until the query
SAVE?“ appears.
Confirm query by pressing the PRINT-key or reject it
by pressing the ESC-key.
Setup
!
Saveing...
Setup29.01.08
02 External adjustment using freely selectable weight user Cal
To select “02 User Cal.“, use the arrow keys ( )
Press -key, on the display appears the weight
value of the adjusting weight. The first digit of the
displayed value is flashing. Now you can make
changes pressing the arrow keys.
To change the number value, press the keys.
To select a digit to the left/right, press the
keys.
Confirm the entered weight value by pressing the
PRINT-key.
Setup29.12.04 Setup