Kern-sohn MLS 50-3C User Manual Page 56

  • Download
  • Add to my manuals
  • Print
  • Page
    / 83
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 55
56 MLS_C-BA-e-1330
To switch the display to the various units of the
measured result, press the M-key.
When drying is finished, you will hear an acoustic
signal and the heating will be shut off.
The display shows the measured result in the set
unit.
Open lid and remove the sample with the help of
the removal aid.
Caution: Caution! Sample dish and all parts of the
sample chamber are hot!
Data output via RS 232
interface is achieved by
pressing the PRINT-key:
--------------- Start drying----------------
Program no. : 1
Name : Floar 100
Profile : Standard
Dry temp. : 120
°
C
Switch off : Automatical 3
Result : Moisture contents -
%M
Printout int : 30 s
Target : None
T. coeficjent : 0.00
Start weight : 9.993 g
0:00:30 0.01 %M
0:01:00 0.02 %M
0:01:30 0.03 %M
Final mass : 9.990 g
Dry time : 0:01:45
Final result : 0.03 %M
--------End of drying procedure------