11.5 Sample material
Easy to determine are usually samples with the following characteristics:
- Grainy to powdery, pourable solids
- Thermally stable materials, emitting the moisture to be determined easily
without other substances evaporating at the same time
- Liquids that vaporize to leave a dry substance without developing a film
Difficult to determine may be samples that are:
- Glutinous or sticky
- Become incrusted easily or tend to form a film
- Decompose easily under the influence of heat or emit various elements
11.6 Sample size / originally weighted in quantity
Drying times, as well as achievable accuracy, are significantly influenced by sample
distribution. In the course of this arise two opposed requirements:
The lighter the originally weighted in quantity, the easier it is to achieve shorter drying
times.
However, the heavier the originally weighted in quantity, the more accurate a result
(example of a sample with ideal characteristics) will be achieved.
Poured
quantity
Reproducibility
0.5g
±0.6%
1g
±0.3%
2g
±0.15%
5g
±0.06%
10g
±0.03%